PHILIPS EPIQ EL18-4
Device
Brand
PHILIPS
Model
EPIQ EL18-4
Serial
****JXBM
Test date
2026-07-14
Customer country
LK
Published
7/14/2026
Repair record
Fault
Lens bubble
Measurements
| Cable | OK, Mid-section coloring of cables |
| Circuit | OK |
| Probe casing | OK |
| Strain relief | OK |
| Matching Layer | OK |
Verify
Scan to verify
https://www.medtrade.world/verification/volka-medical-tech-co-ltd/philips-epiq-el18-4-repair-case
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.
Photos
Fault photos
Pre-repair measurements
| Cable | OK, Mid-section coloring of cables |
|---|---|
| Circuit | OK |
| Probe casing | OK |
| Strain relief | OK |
| Matching Layer | OK |
| Connector casing | OK |
| Crystal detection | Minor damage to the array element, There is interference when touching the probe; the image shows several dark bands, indicating transducer aging. |
| Can test on system | Yes |
| Lens/Membrane/Dome | Bubbling, Damaged |
| Other descriptions | Suggest to replace lens and crystal |
| Had ever been repaired by others? | No |