GE RM6C

Verified
Device
Brand
GE
Model
RM6C
Serial
—
Test date
2020-08-30
Published
8/4/2026
Repair record
Fault

Probe Surveillance Signal (PSS) failed.

Repair Advice

Disassembling the 4D probe for comprehensive testing and chip repair.

Measurements
CableOK
CircuitFaulty
Probe casingOK
Strain reliefOK
Matching LayerOK
Verify
Scan to verify
https://www.medtrade.world/verification/probe-parts/ge-rm6c-repair-case
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.

Photos

Fault photos

Repair process photos

Post-repair photos

Pre-repair measurements

CableOK
CircuitFaulty
Probe casingOK
Strain reliefOK
Matching LayerOK
Connector casingOK
Crystal detectionMinor damage to the array element
Scan image result3D/4D Mode, CW Mode Interference
Can test on systemNo
Lens/Membrane/DomeOK
Other descriptions
Had ever been repaired by others?Unable to judge, Yes