PHILIPS X8-2t
Device
Brand
PHILIPS
Model
X8-2t
Serial
—
Test date
2026-07-10
Published
8/7/2026
Repair record
Fault
No ultrasound image
Repair Advice
No obvious abnormalities were observed in the image after testing on our system; the bending rubber was damaged. Suggested repair solution: Replace the bending rubber.
Measurements
| Cable | OK |
| Circuit | OK |
| Probe casing | OK |
| Strain relief | OK |
| Matching Layer |
Verify
Scan to verify
https://www.medtrade.world/verification/geprobe/philips-x8-2t-repair-case
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.
Photos
Fault photos
Repair process photos
Post-repair photos
Pre-repair measurements
| Cable | OK |
|---|---|
| Circuit | OK |
| Probe casing | OK |
| Strain relief | OK |
| Matching Layer | |
| Connector casing | OK |
| Crystal detection | Minor damage to the array element |
| Scan image result | fault |
| Can test on system | Yes |
| Lens/Membrane/Dome | Damaged |
| Other descriptions | |
| Had ever been repaired by others? | Unable to judge |