Repair record
Fault

No ultrasound image

Repair Advice

No obvious abnormalities were observed in the image after testing on our system; the bending rubber was damaged. Suggested repair solution: Replace the bending rubber.

Measurements
CableOK
CircuitOK
Probe casingOK
Strain reliefOK
Matching Layer
Verify
Scan to verify
https://www.medtrade.world/verification/geprobe/philips-x8-2t-repair-case
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.

Photos

Fault photos

Repair process photos

Post-repair photos

Pre-repair measurements

CableOK
CircuitOK
Probe casingOK
Strain reliefOK
Matching Layer
Connector casingOK
Crystal detectionMinor damage to the array element
Scan image resultfault
Can test on systemYes
Lens/Membrane/DomeDamaged
Other descriptions
Had ever been repaired by others?Unable to judge