ACUSON/SIEMENS Z6MS
Device
Brand
ACUSON/SIEMENS
Model
Z6MS
Serial
—
Test date
2026-07-02
Published
8/7/2026
Repair record
Fault
The image is poor; there's a problem with the leak test.
Repair Advice
Thorough cleaning and replacement of acoustic lens.
Measurements
| Cable | damaged outer skin |
| Circuit | OK |
| Probe casing | |
| Strain relief | OK |
| Matching Layer | OK |
Verify
Scan to verify
https://www.medtrade.world/verification/geprobe/acuson-siemens-z6ms-repair-case
Buyers scan the QR or paste the URL into medtrade.world/v and see the canonical report — matching the printed PDF byte-for-byte.
Photos
Fault photos
Post-repair photos
Pre-repair measurements
| Cable | damaged outer skin |
|---|---|
| Circuit | OK |
| Probe casing | |
| Strain relief | OK |
| Matching Layer | OK |
| Connector casing | OK |
| Crystal detection | OK |
| Scan image result | |
| Can test on system | Yes |
| Lens/Membrane/Dome | Wear and tear |
| Other descriptions | |
| Had ever been repaired by others? | Unable to judge |